Torque screws: Difference between revisions
Jump to navigation
Jump to search
(Created page with "Torque screw force: clockwise force = tightening force counterclockwise force = opening force Medtronic: 0,098 Nm CW and 0,18 Nm CCW Boston Scientific: Orange 0.12 Nm CW and 0....") |
No edit summary |
||
(3 intermediate revisions by the same user not shown) | |||
Line 1: | Line 1: | ||
Torque screw force: | Torque screw force: | ||
clockwise force = tightening force | *clockwise force = tightening force | ||
counterclockwise force = opening force | *counterclockwise force = opening force | ||
Medtronic: | *'''Medtronic: ''' 0.098 Nm CW and 0.18 Nm CCW | ||
Boston Scientific: Orange 0.12 Nm CW and 0.16 Nm CCW | *'''Boston Scientific:''' Orange 0.12 Nm CW and 0.16 Nm CCW | ||
Boston Scientific: Green 0 | *'''Boston Scientific''': Green 0.13 Nm CW and 0.15 Nm CCW | ||
SJM | *'''SJM''': 0.078 Nm CW and CCW |
Latest revision as of 18:29, 16 August 2011
Torque screw force:
- clockwise force = tightening force
- counterclockwise force = opening force
- Medtronic: 0.098 Nm CW and 0.18 Nm CCW
- Boston Scientific: Orange 0.12 Nm CW and 0.16 Nm CCW
- Boston Scientific: Green 0.13 Nm CW and 0.15 Nm CCW
- SJM: 0.078 Nm CW and CCW